RecallTracker
Entrapment hazard~About 550 units affected

Fortemotus Direct Adult Portable Bed Rails Recalled Due to Risk of Serious Injury or Death from Entrapment and Asphyxiation; Violate Mandatory Standard for Adult Portable Bed Rails

What to do (remedy: refund)

Consumers should immediately stop using the recalled adult portable bed rails and contact Fortemotus Direct US for a full refund. Consumers should destroy the bed rails by cutting the handrails' foam padding, take a photo of the destroyed rails and send it to fortemotusofficial@outlook.com

Contact: Fortemotus Direct US by email at fortemotusofficial@outlook.com.

Why it was recalled

The recalled bed rails violate the mandatory standard for adult portable bed rails because users can become entrapped within the bed rail or between the bed rail and the side of the mattress, posing a serious entrapment hazard and risk of death by asphyxiation. The bed rails do not meet structural stability or retention strap requirements, posing a fall hazard. The bed rails' push pins and push pin holes are incorrectly sized, posing a laceration hazard.

Incidents/injuries: None reported

Which products are affected

This recall involves Fortemotus-branded adult portable bed rails, model USFTMTJAS008. The recalled bed rails have black tubing with black foam rubber handle grips, support legs and a fabric pouch, and measure about 30 inches wide by 30 inches high. The Foremotus logo is printed on the bed rail's fabric cover and the model number on the owner's manual.

Photos of the recalled product

Recalled Fortemotus Adult Portable Bed Rail

Recall details

Recall number
26263
Recall date
February 12, 2026
Recalling firm
Fortemotus Direct US
Products
Fortemotus Adult Portable Bed Rails
Sold at
Online at Amazon.com from August 2025 through October 2025 for about $100.
Manufactured in
China
Manufacturers
Guangzhou Kezhinuo Maoyi Youxian Gongsi, dba Fortemotus Direct US, of China
Official notice
View on CPSC.gov →